Long-term stability evaluation of user-derived electrometer calibration factors (kelec, user) for radiotherapy dosimetry.
| Τίτλος: | Long-term stability evaluation of user-derived electrometer calibration factors (k |
|---|---|
| Συγγραφείς: | Tsuno H; Gunma Prefectural College of Health Sciences, Maebashi, Gunma, JAPAN.; Department of Radiation Oncology, Japanese Red Cross Maebashi Hospital, Maebashi, Gunma, JAPAN., Sasaki K; Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, Gunma, JAPAN., Sasaki R; Department of Radiology, National Hospital Organization Sendai Medical Center, Sendai, Miyagi, JAPAN., Matsubayashi F; Department of Radiation Oncology, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Tokyo, JAPAN., Nishi K; Department of Radiology, JR Tokyo General Hospital, Tokyo, JAPAN., Ushikawa T; Department of Radiology, JR Tokyo General Hospital, Tokyo, JAPAN., Goto D; Department of Radiology, JR Tokyo General Hospital, Tokyo, JAPAN., Kosugi R; Department of Radiology, Hamamatsu University School of Medicine, Hamamatsu, Shizuoka, JAPAN., Kawashima Y; Department of Medical Technology, Japanese Red Cross Maebashi Hospital, Maebashi, Gunma, JAPAN. |
| Πηγή: | Journal of applied clinical medical physics [J Appl Clin Med Phys] 2026 Oct; Vol. 27 (10), pp. e70840. |
| Τύπος έκδοσης: | Journal Article; Evaluation Study |
| Γλώσσα: | English |
| Στοιχεία περιοδικού: | Publisher: Wiley on behalf of American Association of Physicists in Medicine Country of Publication: United States NLM ID: 101089176 Publication Model: Print Cited Medium: Internet ISSN: 1526-9914 (Electronic) Linking ISSN: 15269914 NLM ISO Abbreviation: J Appl Clin Med Phys Subsets: MEDLINE |
| Imprint Name(s): | Publication: 2017- : Malden, MA : Wiley on behalf of American Association of Physicists in Medicine Original Publication: Reston, VA : American College of Medical Physics, c2000- |
| Ιατρικοί όροι (MeSH): | Radiometry*/methods , Radiometry*/instrumentation , Radiometry*/standards , Particle Accelerators*/instrumentation , Quality Assurance, Health Care*/standards , Radiotherapy Planning, Computer-Assisted*/methods , Phantoms, Imaging*, Calibration/standards ; Radiotherapy, Intensity-Modulated/methods ; Humans ; Radiotherapy Dosage ; Uncertainty |
| Περίληψη: | Purpose: User-derived electrometer calibration factors (k Methods: A RAMTEC Solo electrometer was used as the reference electrometer, and three electrometers (UNIDOSwebline, RAMTEC Duo, and Super MAX) were evaluated as electrometers under test. Electrometer sensitivity ratios were measured using three signal input methods: A LINAC-based method, a battery-powered current source (SCG002), and a servo-controlled current source (EMF521R). User-derived calibration factors were calculated from the sensitivity ratios and the calibration coefficient of the reference electrometer. Long-term stability was evaluated from repeated measurements over 28 months using linear regression analysis, and the associated uncertainty component was estimated according to the Guide to the Expression of Uncertainty in Measurement (GUM). Results: The coefficients of variation for repeated measurements were ≤ 0.071% for all methods. The k Conclusions: The influence of signal input method on k (© 2026 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.) |
| References: | Cunningham J, Cohen M, Dutreix A. Determination of Absorbed Dose in a Patient Irradiated by Beams of X‐ or Gamma‐Rays in Radiotherapy Procedures. International Commission on Radiation Units and Measurement; 1976. ICRU Report. International Electrotechnical Commission (IEC). Medical electrical equipment—Dosimeters with ionization chambers as used in radiotherapy. Geneva International Electrotechnical Commission; 2011 IEC 60731 :2011. McEwen M, DeWerd L, Ibbott G, et al. Addendum to the AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon beams. Med Phys. 2014;41(4):041501. doi:10.1118/1.4866223. Morgan AM, Aird EG, Aukett RJ, et al. IPEM guidelines on dosimeter systems for use as transfer instruments between the UK primary dosimetry standards laboratory (NPL) and radiotherapy centres. Phys Med Biol. 2000;45(9):2445‐2457. Blad B, Wendel P, Nilsson P. A simple test device for electrometers. Phys Med Biol. 1998;43(8):2385‐2391. Matsubayashi F, Tsuno H, Matsumoto K, et al. Current generator for precision control of electrometer, precision control system, and calibration method. Japan, 2021: Patent 6989834. Kawashima T, Varnava M, Ozawa S, et al. Measurement and evaluation of electrometer correction factors using a charge generator. J Appl Clin Med Phys. 2023;24(8):e14082. doi:10.1002/acm2.14082. Tsuno H, Matsubayashi F, Sasaki K, Sakai T, Matsumoto K, Takeuchi K. Verification of response uncertainty in electrometers through cross‐comparison with a novel current source: a comparative study with guidelines for electrometers used in radiation therapy dosimeters. Jpn J Med Phys. 2024;44(2):21‐28. doi:10.11323/jjmp.44.2_21. Tsuno H, Matsubayashi F, Sasaki K, Sakai T, Matsumoto K, Takeuchi K. Evaluation on the availability of the electrometer inspection by users using a newly developed current source. Radiol Technol. 2023;79(2):166‐176. doi:10.6009/jjrt.2023‐1216. Japanese. Tsuno H, Sasaki R, Sasaki K, et al. Feasibility study of in‐house second‐channel calibration of dual‐channel electrometers using a battery‐powered current source. J Appl Clin Med Phys Vol. 2026;27(3):e70476. Li G, Xiong J, Sun S, et al. Validation of virtual sensor‐assisted Bayesian inference‐based in‐situ sensor calibration strategy for building HVAC systems. Build Simul. 2023;16(2):185‐203. doi:10.1007/s12273‐022‐0935‐7. Deng H, Wang J, Zhang H, Zhang J, Ma M, Zhong X. An in‐situ self‐calibration method for non‐contact full‐field strain measurement. Measurement. 2020;162:107871. doi:10.1016/j.measurement.2020.107871. Koo J, Yoon S, Kim J. Virtual in situ calibration for operational backup virtual sensors in building energy systems. Energies. 2022;15(4):1394. doi:10.3390/en15041394. Joint Committee for Guides in Metrology (JCGM). Evaluation of Measurement Data—Guide to the Expression of Uncertainty in Measurement (GUM). JCGM 100:2008. JCGM; 2008. doi:10.59161/JCGM100‐2008E. Tsuno H, Sasaki K, Matsubayashi F, et al. New idea for determining the optimal number of measurements in absorbed dose‐to‐water measurement of high‐energy photon beam. Radiat Prot Dosimetry. 2022;199(3):246‐253. doi:10.1093/rpd/ncac262. Tsuno H, Maruyama D, Sasaki K, Matsubayashi F. Optimising the number of measurements for absorbed dose‐to‐water calibration coefficient of ionization chamber by cross‐calibration using high‐energy electron beams. J Radioanal Nucl Chem. 2026;335:4335‐4341. doi:10.1007/s10967‐026‐10921‐0. Castro P, García‐Vicente F, Mínguez C, et al. Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration. J Appl Clin Med Phys. 2008;9(1):70‐86. doi:10.1120/jacmp.v9i1.2676. |
| Contributed Indexing: | Keywords: electrometer; electrometer calibration; long‐term stability; quality assurance; radiotherapy dosimetry |
| Entry Date(s): | Date Created: 20261008 Date Completed: 20261008 Latest Revision: 20261010 |