Log file-based patient-specific quality assurance for MR-guided radiation therapy with comprehensive motion management.
| Τίτλος: | Log file-based patient-specific quality assurance for MR-guided radiation therapy with comprehensive motion management. |
|---|---|
| Συγγραφείς: | Yuan K; Medical Physics Division, CUHK Medical Centre, Hong Kong, Hong Kong SAR, China., Cheung MM; Medical Physics Division, CUHK Medical Centre, Hong Kong, Hong Kong SAR, China., Lee L; Medical Physics Division, CUHK Medical Centre, Hong Kong, Hong Kong SAR, China.; Department of Clinical Oncology, Faculty of Medicine, The Chinese University of Hong Kong, Hong Kong, Hong Kong SAR, China. |
| Πηγή: | Journal of applied clinical medical physics [J Appl Clin Med Phys] 2026 Sep; Vol. 27 (9), pp. e70773. |
| Τύπος έκδοσης: | Journal Article |
| Γλώσσα: | English |
| Στοιχεία περιοδικού: | Publisher: Wiley on behalf of American Association of Physicists in Medicine Country of Publication: United States NLM ID: 101089176 Publication Model: Print Cited Medium: Internet ISSN: 1526-9914 (Electronic) Linking ISSN: 15269914 NLM ISO Abbreviation: J Appl Clin Med Phys Subsets: MEDLINE |
| Imprint Name(s): | Publication: 2017- : Malden, MA : Wiley on behalf of American Association of Physicists in Medicine Original Publication: Reston, VA : American College of Medical Physics, c2000- |
| Ιατρικοί όροι (MeSH): | Quality Assurance, Health Care*/standards , Radiotherapy Planning, Computer-Assisted*/methods , Radiotherapy Planning, Computer-Assisted*/standards , Radiotherapy, Image-Guided*/methods , Radiotherapy, Image-Guided*/standards , Particle Accelerators*/instrumentation , Magnetic Resonance Imaging*/methods , Image Processing, Computer-Assisted*/methods , Neoplasms*/radiotherapy, Radiotherapy, Intensity-Modulated/methods ; Humans ; Radiotherapy Dosage ; Movement |
| Περίληψη: | Background: Patient-Specific Quality Assurance (PSQA) for Elekta Unity MR-Linac treatments delivered under Comprehensive Motion Management (CMM) is challenging because beam gating, unexpected interruptions, and split sub-segments introduce delivery complexities that are not fully addressed by conventional quality assurance workflows. Purpose: This study aimed to develop and evaluate a log file-based PSQA program for Elekta Unity MR-Linac treatments delivered under CMM. Methods: Log files were parsed to extract Multileaf Collimator (MLC) positions, jaw positions, and Monitor Units (MU) for each segment, with dedicated handling of CMM-induced beam gating, unexpected beam interruptions, and split sub-segments. Gantry-resolved fluence maps reconstructed from the log files were compared with planned fluence maps using gamma analysis. The system was evaluated for 66 cases, including 3 completion plans and 2 baseline shift plans. Results: Among the 66 plans, 64 achieved passing gamma results, with a pooled average passing rate of 99.91 ± 0.21%. Conclusions: The proposed log file-based PSQA framework demonstrates accuracy and clinical practicality for MR-Linac treatments in CMM mode. It may serve as a reliable post-treatment verification tool and reduce dependence on conventional phantom-based QA. (© 2026 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.) |
| References: | Winkel D, Bol GH, Kroon PS, et al. Adaptive radiotherapy: the Elekta Unity MR‐linac concept. Clin Transl Radiat Oncol. 2019;18:54‐59. doi:10.1016/j.ctro.2019.04.001. Donahue W, Xing S, Smith L, et al. Characterizing the motion management system accuracy on a 1.5T MR‐Linac. Physica Medica. 2025;139:105189. doi:10.1016/j.ejmp.2025.105189. Smith BR, St‐Aubin J, Hyer DE, Commissioning of a motion management system for a 1.5T Elekta Unity MR‐Linac: a single institution experience. J Appl Clin Med Phys. 2025;26(4):e70005. doi:10.1002/acm2.70005. Kabat CN, Defoor DL, Myers P, et al. Evaluation of the Elekta Agility MLC performance using high‐resolution log files. Medical Physics. 2019;46(3):1397‐1407. doi:10.1002/mp.13374. Yuan K, Cheung MM, Lee L, Efficient MLC quality assurance using a virtual picket fence test in an MR‐Linac. J Appl Clin Med Phys. 2025;26(9):e70232. doi:10.1002/acm2.70232. Chow VUY, Kan MWK, Chan ATC, Patient‐specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT). J Appl Clin Med Phys. 2020;21(11):179‐187. doi:10.1002/acm2.13053. Lim SB, Godoy Scripes P, Napolitano M, et al. An investigation of using log‐file analysis for automated patient‐specific quality assurance in MRgRT. J Appl Clin Med Phys. 2021;22(9):183‐188. doi:10.1002/acm2.13361. Menten MJ, Mohajer JK, Nilawar R, et al. Automatic reconstruction of the delivered dose of the day using MR‐linac treatment log files and online MR imaging. Radiother Oncol. 2020;145:88‐94. doi:10.1016/j.radonc.2019.12.010. Viola P, Romano C, Craus M, et al. Prediction of VMAT delivery accuracy using plan modulation complexity score and log‐files analysis. Biomed Phys Eng Express. 2022;8(5):055020. doi:10.1088/2057‐1976/ac82c6. Yuan K, Man‐hin Cheung M, Lee L, Enhancing patient‐specific quality assurance in MR‐guided radiation therapy: a fluence‐based method using log files. J Appl Clin Med Phys. 2025;26(9):e70238. doi:10.1002/acm2.70238. Ahmad SB, Sarfehnia A, Paudel MR, et al. Evaluation of a commercial MRI Linac based Monte Carlo dose calculation algorithm with geant 4. Medical Physics. 2016;43(2):894‐907. doi:10.1118/1.4939808. Huang C‐Y, Yang B, Lam WW, Geng H, Cheung KY, Yu SK, Magnetic field induced dose effects in radiation therapy using MR‐linacs. Medical Physics. 2023;50(6):3623‐3636. doi:10.1002/mp.16397. Yan L, Xu Y, Dai J, Impact of 1.5 T magnetic field on treatment plan quality in MR‐guided radiotherapy: typical phantom test cases. Technol Cancer Res Treat. 2024;23:15330338241272038. doi:10.1177/15330338241272038. Jermoumi M, Xie R, Cao D, Housley DJ, Shepard DM, Does gated beam delivery impact delivery accuracy on an Elekta linac? J Appl Clin Med Phys. 2017;18(1):90‐95. doi:10.1002/acm2.12020. Fujimoto K, Tateoka K, Yaegashi Y, et al. Effects of beam startup characteristics on dose delivery accuracy at low monitor units in step‐and‐shoot intensity modulated radiation therapy. Int J Med Phys Clin Eng Radiat Oncol. 2013;02:1‐5. Klein EE, Hanley J, Bayouth J, et al. Task group 142 report: quality assurance of medical accelerators. Medical Physics. 2009;36(9Part1):4197‐4212. doi:10.1118/1.3190392. Luo G‐w, Deng X‐w, Qi Z‐y, Sun W‐z, Cao W‐f, Investigation on the impact to beam characteristics of a linear accelerator related to duty cycle of respiratory gating. Radiation Measurements. 2011;46(12):1996‐1999. doi:10.1016/j.radmeas.2011.10.003. Hanley J, Dresser S, Simon W, et al. AAPM Task Group 198 Report: an implementation guide for TG 142 quality assurance of medical accelerators. Medical Physics. 2021;48(10):e830‐e885. doi:10.1002/mp.14992. Riis HL, Chick J, Dunlop A, Tilly D, The quality assurance of a 1.5 T MR‐linac. Semin Radiat Oncol. 2024;34(1):120‐128. doi:10.1016/j.semradonc.2023.10.011. Snyder JE, St‐Aubin J, Yaddanapudi S, et al. Commissioning of a 1.5T Elekta Unity MR‐linac: a single institution experience. J Appl Clin Med Phys. 2020;21(7):160‐172. doi:10.1002/acm2.12902. |
| Contributed Indexing: | Keywords: MR‐Linac; comprehensive motion management; log file; patient‐specific quality assurance; radiation therapy |
| Entry Date(s): | Date Created: 20260825 Date Completed: 20260825 Latest Revision: 20260917 |
| Update Code: | 20260918 |
| PubMed Central ID: | PMC13504375 |
| DOI: | 10.1002/acm2.70773 |
| PMID: | 42638481 |
| Βάση Δεδομένων: | MEDLINE |
| ISSN: | 1526-9914 |
|---|---|
| DOI: | 10.1002/acm2.70773 |