Design of a force-measuring setup for colorectal compression anastomosis and first ex-vivo results

Λεπτομέρειες βιβλιογραφικής εγγραφής
Τίτλος: Design of a force-measuring setup for colorectal compression anastomosis and first ex-vivo results
Συγγραφείς: Steger, Jana, Patzke, Isabella, Berlet, Maximilian, Ficht, Stefanie, Eblenkamp, Markus, Mela, Petra, Wilhelm, Dirk
Πηγή: http://lobid.org/resources/99370675073906441#!, 16(8):1335-1345.
Έτος έκδοσης: 2021
Συλλογή: Publisso (ZB MED-Publikationsportal Lebenswissenschaften)
Θεματικοί όροι: Colonoscopy/methods [MeSH], Swine [MeSH], Pressure [MeSH], Rectum/surgery [MeSH], Colon, Colorectal Neoplasms/surgery [MeSH], Biomechanical Phenomena [MeSH], Forces, Colorectal Neoplasms/physiopathology [MeSH], Animals [MeSH], Original Article, Anastomosis, Surgical/methods [MeSH], Implants, Neoplasms, Experimental [MeSH], Digestive System Surgical Procedures/methods [MeSH], Endoscopy, Piercing
Περιγραφή: Purpose!#!The introduction of novel endoscopic instruments is essential to reduce trauma in visceral surgery. However, endoscopic device development is hampered by challenges in respecting the dimensional restrictions, due to the narrow access route, and by achieving adequate force transmission. As the overall goal of our research is the development of a patient adaptable, endoscopic anastomosis manipulator, biomechanical and size-related characterization of gastrointestinal organs are needed to determine technical requirements and thresholds to define functional design and load-compatible dimensioning of devices.!##!Methods!#!We built an experimental setup to measure colon tissue compression piercing forces. We tested 54 parameter sets, including variations of three tissue fixation configurations, three piercing body configurations (four, eight, twelve spikes) and insertion trajectories of constant velocities (5 mms!##!Results!#!There was no statistically significant difference in insertion forces neither between the trajectory groups, nor for variation of tissue fixation configurations. However, we observed a statistically significant increase in insertion forces for increasing number of spikes. The maximum mean peak forces for four, eight and twelve spikes were 6.4 ± 1.5 N, 13.6 ± 1.4 N and 21.7 ± 5.8 N, respectively. The 5th percentile of specimen lumen diameters and pierced tissue thickness were 24.1 mm and 2.8 mm, and the 95th percentiles 40.1 mm and 4.8 mm, respectively.!##!Conclusion!#!The setup enabled reliable biomechanical characterization of colon material, on the base of which design specifications for an endoscopic anastomosis device were derived. The axial implant closure unit must enable axial force transmission of at least 28 N (22 ± 6 N). Implant and applicator diameters must cover a range between 24 and 40 mm, and the implant gap, compressing anastomosed tissue, between 2 and 5 mm.
Τύπος εγγράφου: article in journal/newspaper
Γλώσσα: English
Relation: https://repository.publisso.de/resource/frl:6444539; https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8295116/
DOI: 10.1007/s11548-021-02371-8
Διαθεσιμότητα: https://repository.publisso.de/resource/frl:6444539
https://doi.org/10.1007/s11548-021-02371-8
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8295116/
Rights: https://creativecommons.org/licenses/by/4.0/
Αριθμός Καταχώρησης: edsbas.9519FDFA
Βάση Δεδομένων: BASE
FullText Text:
  Availability: 0
CustomLinks:
  – Url: https://repository.publisso.de/resource/frl:6444539#
    Name: EDS - BASE (ns324271)
    Category: fullText
    Text: View record from BASE
Header DbId: edsbas
DbLabel: BASE
An: edsbas.9519FDFA
RelevancyScore: 906
AccessLevel: 3
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 906.347900390625
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Design of a force-measuring setup for colorectal compression anastomosis and first ex-vivo results
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Steger%2C+Jana%22">Steger, Jana</searchLink><br /><searchLink fieldCode="AR" term="%22Patzke%2C+Isabella%22">Patzke, Isabella</searchLink><br /><searchLink fieldCode="AR" term="%22Berlet%2C+Maximilian%22">Berlet, Maximilian</searchLink><br /><searchLink fieldCode="AR" term="%22Ficht%2C+Stefanie%22">Ficht, Stefanie</searchLink><br /><searchLink fieldCode="AR" term="%22Eblenkamp%2C+Markus%22">Eblenkamp, Markus</searchLink><br /><searchLink fieldCode="AR" term="%22Mela%2C+Petra%22">Mela, Petra</searchLink><br /><searchLink fieldCode="AR" term="%22Wilhelm%2C+Dirk%22">Wilhelm, Dirk</searchLink>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <i>http://lobid.org/resources/99370675073906441#!, 16(8):1335-1345</i>.
– Name: DatePubCY
  Label: Publication Year
  Group: Date
  Data: 2021
– Name: Subset
  Label: Collection
  Group: HoldingsInfo
  Data: Publisso (ZB MED-Publikationsportal Lebenswissenschaften)
– Name: Subject
  Label: Subject Terms
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Colonoscopy%2Fmethods+[MeSH]%22">Colonoscopy/methods [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Swine+[MeSH]%22">Swine [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Pressure+[MeSH]%22">Pressure [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Rectum%2Fsurgery+[MeSH]%22">Rectum/surgery [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Colon%22">Colon</searchLink><br /><searchLink fieldCode="DE" term="%22Colorectal+Neoplasms%2Fsurgery+[MeSH]%22">Colorectal Neoplasms/surgery [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Biomechanical+Phenomena+[MeSH]%22">Biomechanical Phenomena [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Forces%22">Forces</searchLink><br /><searchLink fieldCode="DE" term="%22Colorectal+Neoplasms%2Fphysiopathology+[MeSH]%22">Colorectal Neoplasms/physiopathology [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Animals+[MeSH]%22">Animals [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Original+Article%22">Original Article</searchLink><br /><searchLink fieldCode="DE" term="%22Anastomosis%22">Anastomosis</searchLink><br /><searchLink fieldCode="DE" term="%22Surgical%2Fmethods+[MeSH]%22">Surgical/methods [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Implants%22">Implants</searchLink><br /><searchLink fieldCode="DE" term="%22Neoplasms%22">Neoplasms</searchLink><br /><searchLink fieldCode="DE" term="%22Experimental+[MeSH]%22">Experimental [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Digestive+System+Surgical+Procedures%2Fmethods+[MeSH]%22">Digestive System Surgical Procedures/methods [MeSH]</searchLink><br /><searchLink fieldCode="DE" term="%22Endoscopy%22">Endoscopy</searchLink><br /><searchLink fieldCode="DE" term="%22Piercing%22">Piercing</searchLink>
– Name: Abstract
  Label: Description
  Group: Ab
  Data: Purpose!#!The introduction of novel endoscopic instruments is essential to reduce trauma in visceral surgery. However, endoscopic device development is hampered by challenges in respecting the dimensional restrictions, due to the narrow access route, and by achieving adequate force transmission. As the overall goal of our research is the development of a patient adaptable, endoscopic anastomosis manipulator, biomechanical and size-related characterization of gastrointestinal organs are needed to determine technical requirements and thresholds to define functional design and load-compatible dimensioning of devices.!##!Methods!#!We built an experimental setup to measure colon tissue compression piercing forces. We tested 54 parameter sets, including variations of three tissue fixation configurations, three piercing body configurations (four, eight, twelve spikes) and insertion trajectories of constant velocities (5 mms!##!Results!#!There was no statistically significant difference in insertion forces neither between the trajectory groups, nor for variation of tissue fixation configurations. However, we observed a statistically significant increase in insertion forces for increasing number of spikes. The maximum mean peak forces for four, eight and twelve spikes were 6.4 ± 1.5 N, 13.6 ± 1.4 N and 21.7 ± 5.8 N, respectively. The 5th percentile of specimen lumen diameters and pierced tissue thickness were 24.1 mm and 2.8 mm, and the 95th percentiles 40.1 mm and 4.8 mm, respectively.!##!Conclusion!#!The setup enabled reliable biomechanical characterization of colon material, on the base of which design specifications for an endoscopic anastomosis device were derived. The axial implant closure unit must enable axial force transmission of at least 28 N (22 ± 6 N). Implant and applicator diameters must cover a range between 24 and 40 mm, and the implant gap, compressing anastomosed tissue, between 2 and 5 mm.
– Name: TypeDocument
  Label: Document Type
  Group: TypDoc
  Data: article in journal/newspaper
– Name: Language
  Label: Language
  Group: Lang
  Data: English
– Name: NoteTitleSource
  Label: Relation
  Group: SrcInfo
  Data: https://repository.publisso.de/resource/frl:6444539; https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8295116/
– Name: DOI
  Label: DOI
  Group: ID
  Data: 10.1007/s11548-021-02371-8
– Name: URL
  Label: Availability
  Group: URL
  Data: https://repository.publisso.de/resource/frl:6444539<br />https://doi.org/10.1007/s11548-021-02371-8<br />https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8295116/
– Name: Copyright
  Label: Rights
  Group: Cpyrght
  Data: https://creativecommons.org/licenses/by/4.0/
– Name: AN
  Label: Accession Number
  Group: ID
  Data: edsbas.9519FDFA
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=edsbas&AN=edsbas.9519FDFA
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1007/s11548-021-02371-8
    Languages:
      – Text: English
    Subjects:
      – SubjectFull: Colonoscopy/methods [MeSH]
        Type: general
      – SubjectFull: Swine [MeSH]
        Type: general
      – SubjectFull: Pressure [MeSH]
        Type: general
      – SubjectFull: Rectum/surgery [MeSH]
        Type: general
      – SubjectFull: Colon
        Type: general
      – SubjectFull: Colorectal Neoplasms/surgery [MeSH]
        Type: general
      – SubjectFull: Biomechanical Phenomena [MeSH]
        Type: general
      – SubjectFull: Forces
        Type: general
      – SubjectFull: Colorectal Neoplasms/physiopathology [MeSH]
        Type: general
      – SubjectFull: Animals [MeSH]
        Type: general
      – SubjectFull: Original Article
        Type: general
      – SubjectFull: Anastomosis
        Type: general
      – SubjectFull: Surgical/methods [MeSH]
        Type: general
      – SubjectFull: Implants
        Type: general
      – SubjectFull: Neoplasms
        Type: general
      – SubjectFull: Experimental [MeSH]
        Type: general
      – SubjectFull: Digestive System Surgical Procedures/methods [MeSH]
        Type: general
      – SubjectFull: Endoscopy
        Type: general
      – SubjectFull: Piercing
        Type: general
    Titles:
      – TitleFull: Design of a force-measuring setup for colorectal compression anastomosis and first ex-vivo results
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Steger, Jana
      – PersonEntity:
          Name:
            NameFull: Patzke, Isabella
      – PersonEntity:
          Name:
            NameFull: Berlet, Maximilian
      – PersonEntity:
          Name:
            NameFull: Ficht, Stefanie
      – PersonEntity:
          Name:
            NameFull: Eblenkamp, Markus
      – PersonEntity:
          Name:
            NameFull: Mela, Petra
      – PersonEntity:
          Name:
            NameFull: Wilhelm, Dirk
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 01
              Type: published
              Y: 2021
          Identifiers:
            – Type: issn-locals
              Value: edsbas
            – Type: issn-locals
              Value: edsbas.oa
          Titles:
            – TitleFull: http://lobid.org/resources/99370675073906441#!, 16(8):1335-1345
              Type: main
ResultId 1