Экспрессная оценка и оптимизация предела обнаружения в электротермической атомно-абсорбционной спектрометрии
| Τίτλος: | Экспрессная оценка и оптимизация предела обнаружения в электротермической атомно-абсорбционной спектрометрии |
|---|---|
| Συγγραφείς: | Dyakov, A. O., L'vov, B. V., Novikov, E. V. |
| Στοιχεία εκδότη: | Уральский федеральный университет, 2016. |
| Έτος έκδοσης: | 2016 |
| Θεματικοί όροι: | DETECTION LIMIT, PHOTOMETRIC ERROR, АТОМНО-АБСОРБЦИОННАЯ СПЕКТРОМЕТРИЯ С ЭЛЕКТРОТЕРМИЧЕСКОЙ АТОМИЗАЦИЕЙ, ФОТОМЕТРИЧЕСКАЯ ПОГРЕШНОСТЬ, ATOMIC ABSORPTION SPECTROMETRY, ELECTROTHERMAL ATOMIZATION, ПРЕДЕЛ ОБНАРУЖЕНИЯ |
| Περιγραφή: | The functional dependence of the photometric error from the photomultiplier voltage for atomic absorption spectrometer ZEEnit-650 produced by Analytik Jena, has been experimentally studied. This was used to develop, on the base of the algorithm, proposed by these authors previously, a fast semi-empirical method for estimating the detection limits in measuring the integrated absorbance. It was shown that this dependence, determined for one instrument, is acceptable to the whole series of ZEEnit instruments. The proposed approach is also applicable to other types of spectrometers. The functional dependence of the photometric error takes into account the design features of spectrometers. Some other parameters, as the baseline offset compensation time, integration time, light intensity, the background absorbance, and amplitude and integral values of the non-selective absorption, it is easy to evaluate by the use one-three determinations of the analytical signal. This is sufficient for the calculation and optimization of the detection limits of any analytes. Экспериментально изучена функциональная зависимость фотометрической погрешности атомно-абсорбционного спектрометра ZEEnit-650 фирмы Analytik Jena с графитовым атомизатором от напряжения на фотоумножителе во всем рабочем диапазоне значений. Она использована для проверки корректности быстрого, полуэмпирического способа оценки предела обнаружения интегральной атомной абсорбционности, предложенного авторами ранее. Показано, что такая зависимость, найденная для одного прибора, оказалась приемлемой для всей серии приборов ZEEnit. Предложенный подход применим также к спектрометрам других марок. Предварительно снятая функциональная зависимость фотометрической погрешности учитывает их конструктивные особенности, а остальные параметры, такие как время зануления и интегрирования, амплитудное и интегральное значения неселективной абсорбционности нетрудно определить в ходе одного-трех измерений аналитического сигнала. Этого достаточно для расчета и оптимизации концентрационного и массового пределов обнаружения химических элементов. |
| Τύπος εγγράφου: | Article |
| Περιγραφή αρχείου: | application/pdf |
| Γλώσσα: | Russian |
| DOI: | 10.15826/analitika.2014.18.2.001 |
| Σύνδεσμος πρόσβασης: | http://elar.urfu.ru/handle/10995/42688 |
| Αριθμός Καταχώρησης: | edsair.dedup.wf.002..c0bd63c93c11a484f32e7647c007835d |
| Βάση Δεδομένων: | OpenAIRE |
| FullText | Text: Availability: 0 |
|---|---|
| Header | DbId: edsair DbLabel: OpenAIRE An: edsair.dedup.wf.002..c0bd63c93c11a484f32e7647c007835d RelevancyScore: 788 AccessLevel: 3 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 787.551513671875 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Экспрессная оценка и оптимизация предела обнаружения в электротермической атомно-абсорбционной спектрометрии – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Dyakov%2C+A%2E+O%2E%22">Dyakov, A. O.</searchLink><br /><searchLink fieldCode="AR" term="%22L'vov%2C+B%2E+V%2E%22">L'vov, B. V.</searchLink><br /><searchLink fieldCode="AR" term="%22Novikov%2C+E%2E+V%2E%22">Novikov, E. V.</searchLink> – Name: Publisher Label: Publisher Information Group: PubInfo Data: Уральский федеральный университет, 2016. – Name: DatePubCY Label: Publication Year Group: Date Data: 2016 – Name: Subject Label: Subject Terms Group: Su Data: <searchLink fieldCode="DE" term="%22DETECTION+LIMIT%22">DETECTION LIMIT</searchLink><br /><searchLink fieldCode="DE" term="%22PHOTOMETRIC+ERROR%22">PHOTOMETRIC ERROR</searchLink><br /><searchLink fieldCode="DE" term="%22АТОМНО-АБСОРБЦИОННАЯ+СПЕКТРОМЕТРИЯ+С+ЭЛЕКТРОТЕРМИЧЕСКОЙ+АТОМИЗАЦИЕЙ%22">АТОМНО-АБСОРБЦИОННАЯ СПЕКТРОМЕТРИЯ С ЭЛЕКТРОТЕРМИЧЕСКОЙ АТОМИЗАЦИЕЙ</searchLink><br /><searchLink fieldCode="DE" term="%22ФОТОМЕТРИЧЕСКАЯ+ПОГРЕШНОСТЬ%22">ФОТОМЕТРИЧЕСКАЯ ПОГРЕШНОСТЬ</searchLink><br /><searchLink fieldCode="DE" term="%22ATOMIC+ABSORPTION+SPECTROMETRY%22">ATOMIC ABSORPTION SPECTROMETRY</searchLink><br /><searchLink fieldCode="DE" term="%22ELECTROTHERMAL+ATOMIZATION%22">ELECTROTHERMAL ATOMIZATION</searchLink><br /><searchLink fieldCode="DE" term="%22ПРЕДЕЛ+ОБНАРУЖЕНИЯ%22">ПРЕДЕЛ ОБНАРУЖЕНИЯ</searchLink> – Name: Abstract Label: Description Group: Ab Data: The functional dependence of the photometric error from the photomultiplier voltage for atomic absorption spectrometer ZEEnit-650 produced by Analytik Jena, has been experimentally studied. This was used to develop, on the base of the algorithm, proposed by these authors previously, a fast semi-empirical method for estimating the detection limits in measuring the integrated absorbance. It was shown that this dependence, determined for one instrument, is acceptable to the whole series of ZEEnit instruments. The proposed approach is also applicable to other types of spectrometers. The functional dependence of the photometric error takes into account the design features of spectrometers. Some other parameters, as the baseline offset compensation time, integration time, light intensity, the background absorbance, and amplitude and integral values of the non-selective absorption, it is easy to evaluate by the use one-three determinations of the analytical signal. This is sufficient for the calculation and optimization of the detection limits of any analytes.<br />Экспериментально изучена функциональная зависимость фотометрической погрешности атомно-абсорбционного спектрометра ZEEnit-650 фирмы Analytik Jena с графитовым атомизатором от напряжения на фотоумножителе во всем рабочем диапазоне значений. Она использована для проверки корректности быстрого, полуэмпирического способа оценки предела обнаружения интегральной атомной абсорбционности, предложенного авторами ранее. Показано, что такая зависимость, найденная для одного прибора, оказалась приемлемой для всей серии приборов ZEEnit. Предложенный подход применим также к спектрометрам других марок. Предварительно снятая функциональная зависимость фотометрической погрешности учитывает их конструктивные особенности, а остальные параметры, такие как время зануления и интегрирования, амплитудное и интегральное значения неселективной абсорбционности нетрудно определить в ходе одного-трех измерений аналитического сигнала. Этого достаточно для расчета и оптимизации концентрационного и массового пределов обнаружения химических элементов. – Name: TypeDocument Label: Document Type Group: TypDoc Data: Article – Name: Format Label: File Description Group: SrcInfo Data: application/pdf – Name: Language Label: Language Group: Lang Data: Russian – Name: DOI Label: DOI Group: ID Data: 10.15826/analitika.2014.18.2.001 – Name: URL Label: Access URL Group: URL Data: <link linkTarget="URL" linkTerm="http://elar.urfu.ru/handle/10995/42688" linkWindow="_blank">http://elar.urfu.ru/handle/10995/42688</link> – Name: AN Label: Accession Number Group: ID Data: edsair.dedup.wf.002..c0bd63c93c11a484f32e7647c007835d |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=edsair&AN=edsair.dedup.wf.002..c0bd63c93c11a484f32e7647c007835d |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.15826/analitika.2014.18.2.001 Languages: – Text: Russian Subjects: – SubjectFull: DETECTION LIMIT Type: general – SubjectFull: PHOTOMETRIC ERROR Type: general – SubjectFull: АТОМНО-АБСОРБЦИОННАЯ СПЕКТРОМЕТРИЯ С ЭЛЕКТРОТЕРМИЧЕСКОЙ АТОМИЗАЦИЕЙ Type: general – SubjectFull: ФОТОМЕТРИЧЕСКАЯ ПОГРЕШНОСТЬ Type: general – SubjectFull: ATOMIC ABSORPTION SPECTROMETRY Type: general – SubjectFull: ELECTROTHERMAL ATOMIZATION Type: general – SubjectFull: ПРЕДЕЛ ОБНАРУЖЕНИЯ Type: general Titles: – TitleFull: Экспрессная оценка и оптимизация предела обнаружения в электротермической атомно-абсорбционной спектрометрии Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Dyakov, A. O. – PersonEntity: Name: NameFull: L'vov, B. V. – PersonEntity: Name: NameFull: Novikov, E. V. IsPartOfRelationships: – BibEntity: Dates: – D: 20 M: 11 Type: published Y: 2016 Identifiers: – Type: issn-locals Value: edsair |
| ResultId | 1 |