Academic Journal

Development and Testing of a Tritium -Decay by Means of Electron Integral Detector Prototype Based on Faraday Cup.

Λεπτομέρειες βιβλιογραφικής εγγραφής
Τίτλος: Development and Testing of a Tritium -Decay by Means of Electron Integral Detector Prototype Based on Faraday Cup.
Συγγραφείς: Chernov, V., Berlev, A., Parfenov, V., Razin, V., Titov, N., Tkachev, I., Zadorozhny, S.
Πηγή: Physics of Atomic Nuclei; Dec2025, Vol. 88 Issue 6, p1196-1200, 5p
Θεματικοί όροι: Tritium, Detectors, Sterile neutrinos, Particle detectors, Noise, Spectrum analysis instruments
Περίληψη: To increase the statistical sensitivity of the search for sterile neutrinos in the Tritium -decay, it is proposed to use an integral detector based on a Faraday cup. Detector prototype was designed and tested. The noise level turned to be 2...3 times larger than Schottky limit. Switch time between HV points turned to be less than 4 s. Tritium spectrum shape measured by Faraday cup detector looks quite similar to that measured by standard silicon detector. [ABSTRACT FROM AUTHOR]
Copyright of Physics of Atomic Nuclei is the property of Springer Nature and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)