Test-bench experimental validation of pressure accuracy in a mechanical threshold device for inspiratory muscle training: The threshold IMT® case.
| Τίτλος: | Test-bench experimental validation of pressure accuracy in a mechanical threshold device for inspiratory muscle training: The threshold IMT® case. |
|---|---|
| Συγγραφείς: | Foschiani M; Department of Medicine, University of Udine, Udine, Italy., Bortolotti E; Advantech TIME Srl, Udine, Italy., Chiavone A; Advantech TIME Srl, Udine, Italy., Salvitti S; Department of Medicine, University of Udine, Udine, Italy; Unit of Cardiorespiratory Physiotherapy, Azienda Sanitaria Universitaria Friuli Centrale (ASUFC), Udine, Italy. Electronic address: simone.salvitti@uniud.it. |
| Πηγή: | Respiratory medicine [Respir Med] 2026 Sep; Vol. 261, pp. 109025. Date of Electronic Publication: 2026 Jul 11. |
| Τύπος έκδοσης: | Journal Article; Validation Study |
| Γλώσσα: | English |
| Στοιχεία περιοδικού: | Publisher: Elsevier Country of Publication: England NLM ID: 8908438 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1532-3064 (Electronic) Linking ISSN: 09546111 NLM ISO Abbreviation: Respir Med Subsets: MEDLINE |
| Imprint Name(s): | Publication: 2003- : Oxford : Elsevier Original Publication: London : Baillière Tindall, in association with the British Thoracic Society, [c1989- |
| Ιατρικοί όροι (MeSH): | Respiratory Muscles*/physiology , Inhalation*/physiology , Breathing Exercises*/instrumentation , Breathing Exercises*/methods, Humans ; Reproducibility of Results ; Pressure ; Equipment Design ; Calibration |
| Περίληψη: | Inspiratory muscle training (IMT) is widely used in rehabilitation, yet the accuracy of mechanical threshold devices has been only partially characterised. The aim of this study was to evaluate the validity and stability of the pressure delivered by the Threshold IMT® device (Philips Respironics) using independent calibrated instrumentation, with the measurement chain itself validated against a weighted-plunger reference method. Three units were tested on a bench equipped with a calibrated differential pressure transducer (YOKOGAWA EJA110E; full-scale (FS) error ≤0.055%), an ISO 5167-3-referenced differential flow meter (uncertainty <5%), and an industrial fan. Five measurements per nominal pressure level (9, 25, and 41 cmH (Copyright © 2026 The Authors. Published by Elsevier Ltd.. All rights reserved.) |
| Competing Interests: | Declaration of competing interest Employment: No author is employed by the manufacturer of the device under test (Koninklijke Philips N.V.) or by any entity with a commercial interest in the outcomes of this study. Advantech TIME Srl is a company specializing in fluid dynamics, thermodynamics, and acoustics, with an in-house testing laboratory. The company provided the test bench infrastructure for this study. It has no commercial relationship with Koninklijke Philips N.V. and no financial interest in the outcomes of the pressure accuracy assessment reported herein. Consultancies: No author has served as a consultant for any organisation with a financial interest in the subject matter of this manuscript. Stock ownership: No author holds stocks, shares, or equity in any company with a financial interest in the subject matter of this manuscript. Honoraria: No author has received honoraria related to the subject matter of this manuscript. Paid expert testimony: No author has provided paid expert testimony related to the subject matter of this manuscript. Patent applications or registrations: No author holds or has applied for patents related to the subject matter of this manuscript. Grants or any other funding: This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors. Affiliation with the journal: No author serves or has served as an Editor, Associate Editor, or Advisory Board Member of Respiratory Medicine. |
| Contributed Indexing: | Keywords: Device validation; Inspiratory muscle training; Pressure measurement; Respiratory rehabilitation; Threshold IMT; Threshold pressure device |
| Entry Date(s): | Date Created: 20260711 Date Completed: 20260812 Latest Revision: 20260812 |