Evaluation of Serum Kappa Free Light Chain Reference Interval Using a New Stabilized Calibrator.

Bibliographic Details
Title: Evaluation of Serum Kappa Free Light Chain Reference Interval Using a New Stabilized Calibrator.
Authors: Waggoner D; Department of Pathology and Laboratory Medicine, University of Miami Miller School of Medicine, Miami, FL, United States., Thoren K; Department of Pathology and Laboratory Medicine, University of Miami Miller School of Medicine, Miami, FL, United States., Maynard R; Department of Pathology and Laboratory Medicine, University of Kentucky, Lexington, KY, United States., Twum K; Department of Pathology and Laboratory Medicine, Boston University School of Medicine, Boston, MA, United States., Korpi-Steiner N; Department of Pathology and Laboratory Medicine, University of North Carolina at Chapel Hill, Chapel Hill, NC, United States.
Source: The journal of applied laboratory medicine [J Appl Lab Med] 2026 May 05; Vol. 11 (3), pp. 599-605.
Publication Type: Journal Article
Language: English
Journal Info: Publisher: Oxford University Press Country of Publication: England NLM ID: 101693884 Publication Model: Print Cited Medium: Print ISSN: 2576-9456 (Print) Linking ISSN: 24757241 NLM ISO Abbreviation: J Appl Lab Med Subsets: MEDLINE
Imprint Name(s): Publication: 2020- : Oxford : Oxford University Press
Original Publication: [Washington, DC] : American Association for Clinical Chemistry, [2016]-[2019]
MeSH Terms: Immunoglobulin kappa-Chains*/blood , Immunoglobulin lambda-Chains*/blood, Calibration/standards ; Electrophoresis, Capillary/standards ; Humans ; Reference Values ; Female ; Reproducibility of Results ; Adult
Abstract: Background: Measurement of serum free light chains (FLC) including kappa FLC, lambda FLC, and calculated kappa to lambda FLC ratio play a vital role in the diagnosis, prognosis, and monitoring of plasma cell disorders. Recent concerns regarding upward drift in The Binding Site's kappa FLC measurements have prompted a reformulation of the kappa FLC calibrator by the manufacturer. This study aims to assess the impact of the reformulated/new kappa FLC calibrator on both the kappa FLC reference interval (RI) and the FLC ratio diagnostic range.
Methods: Healthy volunteers (n = 124) from 3 tertiary care medical centers had FLCs measured using both the new and the prior lot of kappa FLC calibrators on an Optilite analyzer (The Binding Site). All samples were screened for monoclonal proteins using capillary electrophoresis. Creatinine was measured to confirm renal function.
Results: Comparison of kappa FLC values from the new vs prior calibrator demonstrated minimal differences. Neither calibrator was able to verify manufacturer-suggested RI in our study population. The proportion of samples that had FLC ratios within the manufacturer's claimed diagnostic range was equivalent between calibrators, with roughly 6% to 7% of study participants falling above the diagnostic range of 0.26 to 1.65.
Conclusion: The new reformulated kappa FLC calibrator from The Binding Site does not support verification of the manufacturer's claimed kappa FLC RI or improve the number of healthy patients falling within the FLC ratio diagnostic range. Both the kappa FLC RI and diagnostic range for FLC ratio should be reevaluated and updated in consensus guidelines.
(© The Author(s) 2026. Published by Oxford University Press on behalf of Association for Diagnostics & Laboratory Medicine.)
Grant Information: Binding Site
Substance Nomenclature: 0 (Immunoglobulin kappa-Chains)
0 (Immunoglobulin lambda-Chains)
Entry Date(s): Date Created: 20260116 Date Completed: 20260626 Latest Revision: 20260626
Update Code: 20260627
DOI: 10.1093/jalm/jfaf208
PMID: 41543034
Database: MEDLINE
Description
ISSN:2576-9456
DOI:10.1093/jalm/jfaf208