Cleaning of photostimulable phosphor plates with chlorhexidine-based substances does not affect image quality of radiographs.
| Τίτλος: | Cleaning of photostimulable phosphor plates with chlorhexidine-based substances does not affect image quality of radiographs. |
|---|---|
| Συγγραφείς: | Borge CP; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba 13414-903, SP, Brazil., Correia-Dantas IF; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba 13414-903, SP, Brazil., Ruiz DC; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba 13414-903, SP, Brazil., Gonzalez-Passos T; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba 13414-903, SP, Brazil., Freitas DQ; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba 13414-903, SP, Brazil., Farias-Gomes A; Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba 13414-903, SP, Brazil. |
| Πηγή: | Dento maxillo facial radiology [Dentomaxillofac Radiol] 2026 May 01; Vol. 55 (4), pp. 362-369. |
| Τύπος έκδοσης: | Journal Article |
| Γλώσσα: | English |
| Στοιχεία περιοδικού: | Publisher: Oxford University Press Country of Publication: England NLM ID: 7609576 Publication Model: Print Cited Medium: Internet ISSN: 1476-542X (Electronic) Linking ISSN: 0250832X NLM ISO Abbreviation: Dentomaxillofac Radiol Subsets: MEDLINE |
| Imprint Name(s): | Publication: January 2024- : [Oxford] : Oxford University Press Original Publication: Erlangen, Germany : University Press Erlangen |
| Ιατρικοί όροι (MeSH): | Chlorhexidine*/chemistry , Radiography, Dental, Digital*/instrumentation , X-Ray Intensifying Screens*, Humans |
| Περίληψη: | Objectives: To evaluate image quality of radiographs obtained with photostimulable phosphor (PSP) plates after cleaning with 2% chlorhexidine. Methods: Eight PSP plates were used: 4 from Express system and 4 from VistaScan system. Two plates from each system were wiped with 2% chlorhexidine in aqueous solution (CHX AQ), and the other 2 plates from each system with 2% chlorhexidine in alcoholic solution (CHX AL). Before cleaning and after every 5 cleaning cycles (up to 100 cleanings), 2 radiographs of an acrylic block and 2 of an aluminium step wedge were acquired with each PSP plate. The image's gray values were measured to assess image brightness, noise, uniformity, and contrast. Two-way repeated measures ANOVA followed by Sidak's post hoc test analysed the influence of the number of cleanings and substance type on these parameters (α = 0.05). Results: Considering the number of cleanings, image brightness varied for CHX AL in the Express system and CHX AQ in the VistaScan system, with no consistent pattern (P < .05). Noise varied for both substances in the VistaScan system, also without a clear pattern (P < .05). In the Express system, regarding the substance type, CHX AL resulted in lower brightness and CHX AQ in lower noise after 75 cleanings (P < .05). Contrast and uniformity were not affected by either the number of cleanings or the substance type (P > .05). Despite numerical variations in some objective parameters, overall image quality remained stable across all cleanings for both substances. No artefacts were observed in the images. Conclusion: Regarding image quality, 2% chlorhexidine-whether aqueous or alcohol-based-can be recommended for PSP plates cleaning. (© The Author(s) 2026. Published by Oxford University Press on behalf of the British Institute of Radiology and the International Association of Dentomaxillofacial Radiology. All rights reserved. For commercial re-use, please contact reprints@oup.com for reprints and translation rights for reprints. All other permissions can be obtained through our RightsLink service via the Permissions link on the article page on our site—for further information please contact journals.permissions@oup.com.) |
| Grant Information: | Coordenação de Aperfeiçoamento de Pessoal de Nível Superior |
| Contributed Indexing: | Keywords: artefacts; chlorhexidine; dental digital radiography; disinfection |
| Substance Nomenclature: | R4KO0DY52L (Chlorhexidine) |
| Entry Date(s): | Date Created: 20260115 Date Completed: 20260626 Latest Revision: 20260626 |
| Update Code: | 20260626 |
| DOI: | 10.1093/dmfr/twag004 |
| PMID: | 41537248 |
| Βάση Δεδομένων: | MEDLINE |
καταχωρήστε σχόλιο πρώτοι!