Reliability Evaluation of a Portable Active Noise Control-Integrated Audiometer for Clinical Hearing Assessment.

Λεπτομέρειες βιβλιογραφικής εγγραφής
Τίτλος: Reliability Evaluation of a Portable Active Noise Control-Integrated Audiometer for Clinical Hearing Assessment.
Συγγραφείς: Gao Q; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Ren C; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Fan X; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Wang M; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Liu Y; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., An X; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Dong X; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Wang X; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China.; School of Aerospace Medicine, Center of Clinical Aerospace Medicine, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Zhou L; 989th Hospital of People's Liberation Army Joint Logistic Support Force, Luoyang, Henan Province, China., Qiu J; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Chen J; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China., Zha D; Department of Otolaryngology-Head and Neck Surgery, Xijing Hospital, Air Force Military Medical University, Xi'an, Shaanxi Province, China.
Πηγή: Ear, nose, & throat journal [Ear Nose Throat J] 2026 May; Vol. 105 (5), pp. NP350-NP359. Date of Electronic Publication: 2025 Oct 30.
Τύπος έκδοσης: Journal Article; Evaluation Study
Γλώσσα: English
Στοιχεία περιοδικού: Publisher: SAGE Publications Country of Publication: United States NLM ID: 7701817 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1942-7522 (Electronic) Linking ISSN: 01455613 NLM ISO Abbreviation: Ear Nose Throat J Subsets: MEDLINE
Imprint Name(s): Publication: Thousand Oaks, CA : SAGE Publications
Original Publication: [New York, Insight Pub. Co.]
Ιατρικοί όροι (MeSH): Audiometry, Pure-Tone*/instrumentation , Audiometry, Pure-Tone*/methods , Hearing Loss*/diagnosis, Humans ; Male ; Reproducibility of Results ; Female ; Middle Aged ; Adult ; Prospective Studies ; Auditory Threshold ; Aged ; Young Adult ; Noise ; Adolescent ; Equipment Design
Περίληψη: Purpose: Hearing loss is a major global health challenge, and pure-tone audiometry remains the gold standard for assessing auditory thresholds. However, conventional clinical audiometry requires specialized sound-proof booths, which limit its accessibility in resource-constrained settings. To address this limitation, we developed and evaluated a portable audiometry system, designated as the E200, incorporating active noise control (ANC) technology with integrated supra-aural earphones.
Methods: In this 2-phase study, reference equivalent threshold sound pressure level calibration was first performed on 183 volunteers comparing E200 to conventional clinical pure-tone audiometric testing. Subsequently, we prospectively enrolled 114 participants with diverse otologic pathologies from tertiary otology clinics to evaluate the validity of the E200 audiometric system.
Results: Data from 228 ears were collected from 114 participants, with 7 frequencies tested/ear, resulting in 1596 sets of data. The two one-sided test equivalence analysis demonstrated that all tested frequencies met the equivalence criterion at the ±5 dB margin. Pure-tone average (PTA) was derived from the arithmetic mean of air-conduction thresholds at standard speech frequencies (0.5, 1, 2, and 4 kHz), with hearing loss severity classified according to WHO (2021) criteria: normal (<20 dB hearing level [HL]), mild-to-moderate (20-<50 dB HL), and moderate-to-severe (≥50 dB HL). Both Bland-Altman plots and Cohen's κ statistics demonstrated good agreement with the PTA results.
Conclusions: This ANC-integrated portable audiometer demonstrates promising potential for hearing screening outside conventional sound-proof booths; however, its long-term clinical value and practical applications require further investigation and validation.
Competing Interests: Declaration of Conflicting InterestsThe authors declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: The E200 devices were provided free of charge by Wuhan Douting Technology Co, Ltd for this clinical trial. The authors declare no financial or non-financial competing interests with the company, and the authors declare that they have no competing interests.
Contributed Indexing: Keywords: active noise control (ANC); hearing screening; portable audiometric device; pure-tone audiometry; sound-proof booth
Entry Date(s): Date Created: 20251031 Date Completed: 20260420 Latest Revision: 20260420