Academic Journal

Lower Versus Higher Cost Sound Level Meters for the Calibration of Clinical Sound Pressure Level Measurements.

Λεπτομέρειες βιβλιογραφικής εγγραφής
Τίτλος: Lower Versus Higher Cost Sound Level Meters for the Calibration of Clinical Sound Pressure Level Measurements.
Συγγραφείς: Maryn, Youri1,2,3,4 youri.maryn@zas.be, Sismono, Fergio1
Πηγή: American Journal of Speech-Language Pathology. May2026, Vol. 35 Issue 3, p1268-1273. 6p.
Θεματικοί όροι: *Pearson correlation (Statistics), *Cost analysis, *Audiometry, *Descriptive statistics, *Speech evaluation, *Communication, *Calibration, *Comparative studies, *Data analysis software, *Phonetics, *Transducers, Physiological aspects of speech
Γεωγραφικοί όροι: Belgium
Περίληψη: Purpose: Clinical measurements of voice/speech sound pressure level (SPL) require calibration. However, relatively expensive sound level meter (SLM) may be required, which can be problematic for clinicians with limited resources. Method: This study therefore compared performance (correlation and error) of 11 lower cost mobile communication device (MCD)-based SLMs (SLMMCD; without or with miniature measurement microphones and dedicated software) with Class 2 SPL measurements. Results: Only three SLMMCDs provided accurate SPL measurements. In four other SLMMCDs, response was nonlinear. Conclusion: SPL measurements with low-cost and MCD-based SLMs can be accurate, but, in this study, only for miniature measurement microphones for which factory-determined correction terms were available. [ABSTRACT FROM AUTHOR]